Abstract
Older institutionalized individuals with reduced residual abilities are going to growth in number in the next years. Monitoring the daily energy expenditure may allow a correct dietary intake and an effective physical activity program, thus improving their life quality. Wearable multisensors (WMS) have been developed to estimate energy expenditure (EE), but, if not calibrated, their accuracy may become questionable. This study was aimed at evaluating WMS accuracy in EE estimation in institutionalized elderly people where individual calibration is difficult to perform.

Methods
Twenty-three participants (age 48 ± 27 years; body mass 66.6 ± 12.7 kg; stature 1.69 ± 0.11 m; mean ± SD) were divided in young (YNG, <30 years; n = 9), adult (AD, 30–60 years, n = 7), and institutionalized elderly (ELD, >60 years; n = 7) individuals. EE at two different exercise intensities during cycling with the upper (UA) or lower limbs (LL) and walking was measured by indirect calorimetry (IC) and estimated by WMS.

Results
Estimated EE was always significantly lower than in IC in YNG and AD (P < 0.05). WMS underestimated EE in all conditions and the error increased with physical activity intensity. In YNG and ELD the correlation coefficient between measured and estimated EE improved during walking (R = 0.61 and R = 0.34 in YNG and ELD, respectively). In AD the best correlation occurred during LL (R = 0.78).

Conclusions
Without individual calibration, WMS underestimated EE measured by IC during all physical activities, with the lower correlation coefficient in institutionalized individuals. An individual calibration of WMS seems to be mandatory to obtain accurate estimations of EE, especially in institutionalized elderly people.

Direct Link: https://doi.org/10.1007/s11332-016-0321-x

Journal: Sport Sciences for Health. 2016 Dec 1;12(3):463-70

Keywords: daily metabolic expenditure, metabolism, nursing home, older adults, Physical Activity,

Applications: Physical Activity,

CamNtech Reference: AH16031

Back to Search Results

UK & International customers

CamNtech Ltd.
Manor Farm
Fenstanton
Cambridgeshire
PE28 9JD, UK

US customers

CamNtech Inc.
630 Boerne Stage Airfield,
Boerne,
Texas 78006,
USA

Copyright

© 2024 CamNtech Ltd and CamNtech Inc

Company information

Registered in England No. 2221302
VAT No: GB486 3019 34


Privacy Policy