OBJECTIVE To evaluate the safety and efficacy of closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII).

RESEARCH DESIGN AND METHODS A total of 12 women with type 1 diabetes (aged 32.9 years, diabetes duration 17.6 years, BMI 27.1 kg/m2, and HbA1c 6.4%) were randomly allocated to closed-loop or conventional CSII. They performed normal daily activities (standardized meals, snacks, and exercise) for 24 h on two occasions at 19 and 23 weeks’ gestation. Plasma glucose time in target (63–140 mg/dL) and time spent hypoglycemic were calculated.

RESULTS Plasma glucose time in target was comparable for closed-loop and conventional CSII (median [interquartile range]: 81 [59–87] vs. 81% [54–90]; P = 0.75). Less time was spent hypoglycemic (<45 mg/dL [0.0 vs. 0.3%]; P = 0.04), with a lower low blood glucose index (2.4 [0.9–3.5] vs. 3.3 [1.9–5.1]; P = 0.03), during closed-loop insulin delivery.

CONCLUSIONS Closed-loop insulin delivery was as effective as conventional CSII, with less time spent in extreme hypoglycemia.

Direct Link: https://doi.org/10.2337/dc11-1430

Journal: Diabetes Care. 2011 Dec 1;34(12):2527-9

Keywords: diabetes, energy expenditure, insulin delivery, pregnancy,

Applications: Energy Expenditure,

CamNtech Reference: AH11018