Background
The accuracy of physical activity (PA) monitors to discriminate between PA, sedentary behavior, and non-wear in extremely obese (EO) adolescents is unknown.

Methods
Twenty-five subjects (9 male/16 female; age=16.5±2.0 y; BMI=51±8 kg/m2) wore three activity monitors (StepWatch [SAM], Actical [AC], Actiheart [AH]) during a 400 meter walk test (400MWT), two standardized PA bouts of varying duration, and one sedentary bout.

Results
For the 400MWT, percent error between observed and monitor recorded steps was 5.5±7.1% and 82.1±38.6% for the SAM and AC steps, respectively (observed vs. SAM steps: −17.2±22.2 steps; observed vs. AC steps: −264.5±124.8 steps). All activity monitors were able to differentiate between PA and sedentary bouts but only SAM steps and AH heart rate were significantly different between sedentary behavior and non-wear (p<0.001). For all monitors, sedentary behavior was characterized by bouts of zero steps/counts punctuated by intermittent activity steps/counts; non-wear was represented almost exclusively by zero steps/counts.

Conclusion
Of all monitors tested, the SAM was most accurate in terms of counting steps and differentiating levels of PA, and thus, most appropriate for EO adolescents. The ability to accurately characterize PA intensity in EO adolescents critically depends on activity monitor selection.

Direct Link: https://www.ncbi.nlm.nih.gov/pubmed/25205688

Journal: Journal of physical activity & health. 2015 Jan;12(1):132

Keywords: adolescents, obesity, Physical Activity, Sedentary behaviour, teenagers, validation,

Applications: Physical Activity,

CamNtech Reference: AH15024

Back to Search Results

UK & International customers

CamNtech Ltd.
Manor Farm
Fenstanton
Cambridgeshire
PE28 9JD, UK

US customers

CamNtech Inc.
630 Boerne Stage Airfield,
Boerne,
Texas 78006,
USA

Copyright

© 2024 CamNtech Ltd and CamNtech Inc

Company information

Registered in England No. 2221302
VAT No: GB486 3019 34


Privacy Policy