Objective
A number of small portable systems that can measure HRV are available to address questions related to autonomic regulation in free-living subjects. However, ambulatory HRV measurements obtained through use of these systems have not previously been validated against standard clinical measurements such as Holter recordings. The objective of this study was to validate HRV obtained using a commonly used system, Actiheart, during occupational and leisure-time activities.

Method
Full-day ambulatory electrocardiography (ECG) signals were recorded from 8 females simultaneously using Actiheart and Holter recorders, and signals were processed to RR-interval time series. Segments of 5-minute duration were sampled every 30 minutes, and spectral components of the heart rate variability were calculated. Actiheart and Holter values were compared using Deming regression analysis and Bland-Altman plots.

Results
In total, 489 segments were available with an HRV value from both Actiheart and Holter recordings after filtering out segments with >10% interpolated beats. No systematic differences between Actiheart and Holter HRV were found. The random deviations between Actiheart and Holter were comparable to the repeatability standard deviation between consecutive Holter measurements.

Discussion
The results show that Actiheart is suited as a stand-alone ambulatory method for heart rate variability monitoring during occupational and leisure-time activities.

Direct Link: https://doi.org/10.1186/1475-925X-10-27

Journal: . Biomedical engineering online. 2011 Dec;10(1):1-4.

Keywords: autonomic nervous system, comparison, heart rate variability, holter monitoring, validity,

Applications: HRV,

CamNtech Reference: AH11006

Back to Search Results

UK & International customers

CamNtech Ltd.
Manor Farm
Fenstanton
Cambridgeshire
PE28 9JD, UK

US customers

CamNtech Inc.
630 Boerne Stage Airfield,
Boerne,
Texas 78006,
USA

Copyright

© 2020 CamNtech Ltd and CamNtech Inc

Company information

Registered in England No. 2221302
VAT No: GB486 3019 34


Privacy Policy