Submaximal step tests are often used for estimation of maximum oxygen consumption (VO2max) in humans. The validity of the Actiheart step test for VO2

max estimation was not fully studied yet. Therefore, purpose of the study was to estimate VO2 max using the Actiheart step test and to compare the data with measured VO2 max in endurance trained athletes (ATH) and healthy non-athletes (CON). 68 ATH (54% men, 28.0±5.4 yrs, 20.9±1.7 kg∙m-2) and 63 CON (46% men, 27.6±5.1 yrs, 22.1±1.7 kg∙m-2) performed the Actiheart step test and a spiroergometry for assessment of VO2max. In addition, resting metabolic rate (RMR; indirect calorimetry), maximum heart rate (HRmax; heart rate monitoring

system during spiroergometry), and sleeping heart rate (SHR; Actiheart 6-day long term measurement) were determined. Validity of two different Actiheart software entry modes was assessed: (1) AHraw (estimated RMR [Schofield] and HRmax [Tanaka], SHR = 70 bpm) and (2) AHcomplete (measured RMR, HRmax, and SHR). Validity was investigated using linear regression (R2 and standard error of the estimate (SEE)) and repeated-measures ANOVA with a Bonferroni post-hoc correction. The level of significance was set to α=0.05. VO2max estimated by AHraw was significant related to measured VO2max in women CON (R2=0.22;

p<0.05), whereas when VO2max was estimated by AHcomplete the relation was significant in women ATH and CON, and in men CON (R2=0.17-0.24; p<0.05). AHraw significantly underestimated VO2max in the total sample by 8% (51.4±10.2 vs. 55.9±7.6 ml∙kg-1∙min-1; p<0.0001), whereas no significant difference between AHcomplete and the criterion method was found (57.0±11.1 vs. 55.9±7.6 ml∙

kg-1∙min-1; p=0.26). The Actiheart step test is an acceptable tool for the estimation of VO2max if an error within 8% can be tolerated. However, accuracy of the VO2

max prediction is much improved when entering measured variables, such as RMR, SHR, and HRmax, into the software.

Direct Link: https://serval.unil.ch/resource/serval:BIB_A40CD2797772.P001/REF.pdf

Journal: Current Issues in Sport Science 4:003.

Keywords: athletes, cardiorespiratory fitness, exercise testing, Maximum oxygen consumption, metabolic equivalent, sport, validity,

Applications: Energy Expenditure,

CamNtech Reference: AH19002

Back to Search Results

UK & International customers

CamNtech Ltd.
Manor Farm
Fenstanton
Cambridgeshire
PE28 9JD, UK

US customers

CamNtech Inc.
630 Boerne Stage Airfield,
Boerne,
Texas 78006,
USA

Copyright

© 2024 CamNtech Ltd and CamNtech Inc

Company information

Registered in England No. 2221302
VAT No: GB486 3019 34


Privacy Policy