Background/Objective:
The Actiheart (Mini Mitter, Sunriver, OR, USA) uses heart rate (HR) and activity data to predict activity energy expenditure (AEE). Currently, the Actiheart has only been tested during laboratory conditions. Therefore, the objective of this study was to validate the Actiheart prediction method against indirect calorimetry during a wide range of activities in a field setting.

Subjects/Methods:
Forty-eight participants (age: 35±11.4 years) were recruited for the study. Eighteen activities were split into three routines of six activities and each routine was performed by 20 participants. During each routine, the participants wore an Actiheart and simultaneously, AEE was measured with a Cosmed K4b2 portable metabolic system. The manufacturer’s HR algorithm, activity algorithm, and combined activity and HR algorithm were used to estimate AEE.

Results:
The mean error (and 95% prediction intervals) for the combined activity and HR algorithm, HR algorithm, and activity algorithm versus the Cosmed K4b2 were 0.02 kJ kg−1 min−1 (−0.17, 0.22 kJ kg−1 min−1), −0.03 kJ kg−1 min−1 (−0.24, 0.18 kJ kg−1 min−1), and 0.14 kJ kg−1 min−1 (−0.12, 0.40 kJ kg−1 min−1), respectively.

Conclusion:
The Actiheart combined activity and HR algorithm and HR algorithm provide similar estimates of AEE on both a group and individual basis.

Direct Link: https://doi.org/10.1038/sj.ejcn.1602766

Journal: European journal of clinical nutrition. 2008 Jun;62(6):704-11.

Keywords: Heart Rate, oxygen consumption, Physical Activity, validity,

Applications: Energy Expenditure,

CamNtech Reference: AH07004

Back to Search Results

UK & International customers

CamNtech Ltd.
Manor Farm
Fenstanton
Cambridgeshire
PE28 9JD, UK

US customers

CamNtech Inc.
630 Boerne Stage Airfield,
Boerne,
Texas 78006,
USA

Copyright

© 2024 CamNtech Ltd and CamNtech Inc

Company information

Registered in England No. 2221302
VAT No: GB486 3019 34


Privacy Policy